■ functional features
▲ the instrument can measure the single capacitance of groups of shunt capacitors (single-phase capacitance and three-phase capacitance can be measured) without disconnection. At the same time, the instrument can also measure the inductance of various reactors to meet a variety of on-site uses. The three-phase capacitance measurement is connected at one time according to the connection mode of the capacitor bank and the wiring diagram. The instrument has the built-in automatic commutation wiring function, which eliminates the cumbersome wiring commutation short circuit operation, and the test results are automatically generated after the test is completed.
▲ during measurement, the instrument can display the measured voltage, current, power, frequency, impedance, phase angle and other data while displaying the measured capacitance value or inductance value, so as to better analyze the quality of the test object.
▲ the instrument adopts 8.0 inch 800 × 600 graphics dot matrix K600 + core 65K color dgus screen, touch operation, clear observation during the day and night, Chinese menu prompt, easy to operate.
▲ the instrument has built-in large capacity nonvolatile memory: it can store 200 groups of measurement data. The instrument is equipped with USB flash disk interface, which can store any group of measurement data (limited by the capacity of USB flash disk).
▲ the instrument has built-in high-precision real-time clock function: date and time calibration can be carried out.
▲ the instrument is equipped with a high-speed micro thermal printer: it can print measurement and historical data.
▲ overcurrent protection function of instrument test power supply: short circuit of power output will not damage the instrument.
■ technical indicators
▲ test voltage: AC 25V ± 10%, 50Hz AC 1.25V ± 10%, 50Hz 25VA
▲ measurement range and accuracy:
Measurable capacitance range: 0.1uF ~ 6800uf ± (Reading 1% + 0.01uF)
Measurable inductance range: 100uh ~ 20h ± (Reading 3% + 0.05uh)
Measurable current range: 5mA ~ 20a ± (Reading 3% + 0.05ma)
Measurable resistance range: 50m Ω ~ 20K Ω ± (Reading 3% + 0.1M Ω)
▲ overall dimension: 405mm × 330mm × 180mm
▲ instrument weight: 8.5Kg (excluding test line)