■ Technical parameters
Input quantity | MV1000: 4, 6, 8, 12 or 24 channels MV2000: 8, 10, 20, 30, 40 or 48 channels |
Measurement period | MV1004、MV1008、MV2008:125 ms,250ms In high-speed mode: 25ms* MV1006、MV1012、MV1024、MV2010、MV2020、MV2030、 MV2040, MV2048: 1s (A/D integration time cannot be 100ms), 2s, 5s, In high-speed mode: 125ms* In high-speed mode, the A/D integration time is fixed at 1.67ms |
Use high-speed mode Points to note when formulating | When MVAdvanced uses high-speed mode (A/D integration time is 1.67ms), the power frequency Rate interference and other factors may cause fluctuations in the measured value. In this case, please use Normal mode for measurement (A/D integration time is 16.7ms, 20ms or 100ms). |